CVE-2014-5431 Hardcoded credentials in SIGMA Spectrum Infusion System

Medium 6.8 Published Updated

Hard-coded credentials in Baxter infusion pumps allow attackers with physical access to modify biomedical settings and disable alarms.

Exploited in the wild
Not known
Login needed
Not needed
Attackable
Local or same network only
Victim action needed
Not needed
Fix
None listed

What is CVE-2014-5431?

CVE-2014-5431 is a medium-severity vulnerability in Baxter SIGMA Spectrum Infusion System, published on 26 Mar 2019. It can only be exploited from the same machine or local network, with no action from the victim.

Official description

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected. The hard-coded password may allow an attacker with physical access to the device to access management functions to make unauthorized configuration changes to biomedical settings such as turn on and off wireless connections and the phase-complete audible alarm that indicates the end of an infusion phase. Baxter has released a new version of the SIGMA Spectrum Infusion System, version 8, which incorporates hardware and software changes.

Weakness: CWE-798, CWE-259

Which versions are affected by CVE-2014-5431?

It affects 1 product.

Products include Baxter Sigma Spectrum Infusion System Firmware.

How do I fix CVE-2014-5431?

No fix has been published for CVE-2014-5431 yet. Limit exposure of the affected software until one is available.

Is CVE-2014-5431 being exploited?

There is no known exploitation of CVE-2014-5431 in the wild and no public exploit code.