CVE-2014-5434 Hardcoded credentials in SIGMA Spectrum Infusion System
An attacker can access the infusion pump's wireless module via FTP using hard-coded default credentials, bypassing authentication controls.
- Exploited in the wild
- Not known
- Login needed
- Not needed
- Attackable
- Over the internet
- Victim action needed
- Not needed
- Fix
- None listed
What is CVE-2014-5434?
CVE-2014-5434 is a critical vulnerability in Baxter SIGMA Spectrum Infusion System, published on 26 Mar 2019. An attacker can exploit it over the internet without logging in, with no action from the victim.
Official description
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 has a default account with hard-coded credentials used with the FTP protocol. Baxter asserts no files can be transferred to or from the WBM using this account. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
Which versions are affected by CVE-2014-5434?
It affects 1 product.
Products include Baxter Sigma Spectrum Infusion System Firmware.
How do I fix CVE-2014-5434?
No fix has been published for CVE-2014-5434 yet. Limit exposure of the affected software until one is available.
Is CVE-2014-5434 being exploited?
There is no known exploitation of CVE-2014-5434 in the wild and no public exploit code.
Other vulnerabilities in Baxter SIGMA Spectrum Infusion System
- CVE-2020-12040 Information disclosure in Baxter Sigma Spectrum Infusion Pumps
- CVE-2020-12043 Security feature bypass in Baxter Sigma Spectrum Infusion Pumps
- CVE-2020-12045 Hardcoded credentials in Baxter Sigma Spectrum Infusion Pumps
- CVE-2020-12047 Hardcoded credentials in Baxter Sigma Spectrum Infusion Pumps
- CVE-2014-5432 Auth bypass in SIGMA Spectrum Infusion System
Sources
Also known as GHSA-8j2v-wpmp-5j96.