CVE-2014-5434 Hardcoded credentials in SIGMA Spectrum Infusion System

Critical 9.8 Published Updated

An attacker can access the infusion pump's wireless module via FTP using hard-coded default credentials, bypassing authentication controls.

Exploited in the wild
Not known
Login needed
Not needed
Attackable
Over the internet
Victim action needed
Not needed
Fix
None listed

What is CVE-2014-5434?

CVE-2014-5434 is a critical vulnerability in Baxter SIGMA Spectrum Infusion System, published on 26 Mar 2019. An attacker can exploit it over the internet without logging in, with no action from the victim.

Official description

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 has a default account with hard-coded credentials used with the FTP protocol. Baxter asserts no files can be transferred to or from the WBM using this account. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.

Weakness: CWE-798, CWE-259

Which versions are affected by CVE-2014-5434?

It affects 1 product.

Products include Baxter Sigma Spectrum Infusion System Firmware.

How do I fix CVE-2014-5434?

No fix has been published for CVE-2014-5434 yet. Limit exposure of the affected software until one is available.

Is CVE-2014-5434 being exploited?

There is no known exploitation of CVE-2014-5434 in the wild and no public exploit code.