CVE-2015-3952 Information disclosure in Plum A+ Infusion System

High 7.5 Published Updated

Attackers can read plaintext wireless keys from Hospira infusion pumps, potentially enabling unauthorized network access or device control.

Exploited in the wild
Not known
Login needed
Not needed
Attackable
Over the internet
Victim action needed
Not needed
Fix
None listed

What is CVE-2015-3952?

CVE-2015-3952 is a high-severity vulnerability in Hospira Plum A+ Infusion System, published on 25 Mar 2019. An attacker can exploit it over the internet without logging in, with no action from the victim.

Official description

Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

Weakness: CWE-200, CWE-312

Which versions are affected by CVE-2015-3952?

It affects 3 products.

Products include Pifzer Plum A\+3 Infusion System Firmware, Pifzer Plum A\+ Infusion System Firmware, Pifzer Symbiq Infusion System Firmware.

How do I fix CVE-2015-3952?

No fix has been published for CVE-2015-3952 yet. Limit exposure of the affected software until one is available.

Is CVE-2015-3952 being exploited?

There is no known exploitation of CVE-2015-3952 in the wild and no public exploit code.