{"answers":{"affected":"It affects 3 products.","exploited":"There is no known exploitation of CVE-2015-3953 in the wild and no public exploit code.","fix":"No fix has been published for CVE-2015-3953 yet. Limit exposure of the affected software until one is available.","what":"CVE-2015-3953 is a critical vulnerability in Hospira Plum A+ Infusion System, published on 25 Mar 2019. An attacker can exploit it over the internet without logging in, with no action from the victim."},"fix_prompt":"Check this codebase for CVE-2015-3953 and fix it if it is affected.\n\nWhat it is: CVE-2015-3953 is a critical vulnerability in Hospira Plum A+ Infusion System, published on 25 Mar 2019. An attacker can exploit it over the internet without logging in, with no action from the victim.\nExploitation: There is no known exploitation of CVE-2015-3953 in the wild and no public exploit code.\n\nAffected product: Hospira Plum A+ Infusion System. No package-level version data is available, so check the vendor advisory linked from the reference page.\n\nTask:\n1. Search the dependency manifests and lockfiles (for example package.json, package-lock.json, requirements.txt, poetry.lock, uv.lock, pom.xml, build.gradle, go.mod, Cargo.lock, Gemfile.lock) and any Dockerfiles for the affected packages, including transitive dependencies.\n2. If an affected version is in use, upgrade it to the fixed version for its release line with the project's package manager, so the lockfile is regenerated. Prefer the smallest upgrade that fixes it.\n3. If there is no fix, or the upgrade would break the build, do not guess: explain the risk and the safest workaround instead.\n4. Run the tests and summarise what changed and why.\n5. If nothing in the project is affected, say so and change nothing.\n\nReference: https://insights.sbomify.com/CVE-2015-3953/ (Markdown: https://insights.sbomify.com/CVE-2015-3953/index.md)","id":"CVE-2015-3953","markdown":"https://insights.sbomify.com/CVE-2015-3953/index.md","name":"CVE-2015-3953","record":{"aliases":["GHSA-vw52-2pgm-8cj2"],"analysis":{"component_type":"ics_ot","fix_type":"upgrade","impacts":["account_takeover","privilege_escalation"],"network_reachable":"yes","non_default_config":false,"product_category":"Infusion pump system","requires_auth":"none","summary":"Hard-coded accounts allow unauthorized access to Hospira infusion systems, requiring users to upgrade to the Plum 360 model or disable specific ports.","user_interaction":"not_required","vuln_class":"hardcoded_credentials"},"analysis_generated_by":"qwen3.8-flash-next","assigner":"icscert","cpe":[{"product":"plum_a\\+3_infusion_system_firmware","vendor":"pifzer"},{"product":"plum_a\\+_infusion_system_firmware","vendor":"pifzer"},{"product":"symbiq_infusion_system_firmware","vendor":"pifzer"}],"cvss":{"score":9.8,"severity":"critical","source":"nvd","vector":"CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H","version":"3.0"},"cwes":[{"id":"CWE-798","name":"Use of Hard-coded Credentials"},{"id":"CWE-259","name":"Use of Hard-coded Password"}],"description":"Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.","ecosystems":[],"exploits":{"exploitdb":0,"first_seen":null,"github_poc":0,"max_poc_stars":null,"metasploit":0,"nuclei":0,"refs":[],"tier":"none_known"},"has_fix":null,"headline":"Hardcoded credentials in Plum A+ Infusion System","id":"CVE-2015-3953","indexable":true,"kev":null,"kind":"cve","links":{"ecosystem":null,"product":null,"products":[],"type":"hardcoded_credentials","year":"2015"},"n_packages":0,"n_references":1,"next":{"headline":"Auth bypass in Plum A+ Infusion System","id":"CVE-2015-3954"},"packages":[],"prev":{"headline":"Information disclosure in Plum A+ Infusion System","id":"CVE-2015-3952"},"products":["Plum A+ Infusion System","Plum A+3 Infusion System","Symbiq Infusion System"],"published":"2019-03-25","related":[{"cvss_score":9.8,"headline":"Auth bypass in Plum A+ Infusion System","id":"CVE-2015-3954","kev_date_added":null,"published":"2019-03-25"},{"cvss_score":9.8,"headline":"Missing authorization in Plum A+ Infusion System","id":"CVE-2015-3956","kev_date_added":null,"published":"2019-03-25"},{"cvss_score":7.5,"headline":"Information disclosure in Plum A+ Infusion System","id":"CVE-2015-3952","kev_date_added":null,"published":"2019-03-25"}],"sources":[{"class":"attribution","licence":"CVE Terms of Use (MITRE)","name":"CVE List","prefix":"CVE"},{"class":"attribution","licence":"US public information, NVD API terms","name":"NVD"},{"class":"attribution","licence":"CWE Terms of Use","name":"CWE (MITRE)"},{"class":"attribution","licence":"FIRST CVSS licence","name":"CVSS (FIRST)"}],"ssvc":{"automatable":null,"exploitation":null,"technical_impact":null},"state":"PUBLISHED","updated":"2024-08-06","vendors":["Hospira"]},"summary":"Hard-coded accounts allow unauthorized access to Hospira infusion systems, requiring users to upgrade to the Plum 360 model or disable specific ports.","url":"https://insights.sbomify.com/CVE-2015-3953/"}