CVE-2017-9657 Business logic in IntelliVue MX40 Patient Worn Monitor

Medium 6.5 Published Updated

Attackers can cause a patient monitor to fail local alarming during network re-association, delaying treatment for critical events.

Exploited in the wild
Not known
Login needed
Not needed
Attackable
Local or same network only
Victim action needed
Not needed
Fix
None listed

What is CVE-2017-9657?

CVE-2017-9657 is a medium-severity vulnerability in Philips IntelliVue MX40 Patient Worn Monitor, published on 30 Apr 2018. It can only be exploited from the same machine or local network, with no action from the victim.

Official description

Under specific 802.11 network conditions, a partial re-association of the Philips IntelliVue MX40 Version B.06.18 WLAN monitor to the central monitoring station is possible. In this state, the central monitoring station can indicate the MX40 is not connected or associated to the central monitor, and thus should be operating in local monitoring mode (local audio-on, screen-on), but the MX40 WLAN itself can instead still be operating in telemetry mode (local audio-off, screen-off). If a patient experiences an alarm event and clinical staff expects the MX40 to provide local alarming when it is not available from the local device, a delay of treatment can occur. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.

Weakness: CWE-755, CWE-460

Which versions are affected by CVE-2017-9657?

It affects 1 product.

Products include Philips Intellivue Mx40 Firmware.

How do I fix CVE-2017-9657?

No fix has been published for CVE-2017-9657 yet. Limit exposure of the affected software until one is available.

Is CVE-2017-9657 being exploited?

There is no known exploitation of CVE-2017-9657 in the wild and no public exploit code.

Sources

Also known as GHSA-32rc-pv7x-94gg.

Is CVE-2017-9657 in your software?

Upload an SBOM and sbomify checks it against every vulnerability here, every day.

Check your SBOMs for free