CVE-2018-10601 Memory corruption in IntelliVue Patient Monitors, Avalon Fetal/Maternal Monitors

High 8.2 Published Updated

An attacker on the same subnet can trigger a stack overflow via the echo service, potentially leading to code execution or denial of service.

Exploited in the wild
Not known
Login needed
Not needed
Attackable
Local or same network only
Victim action needed
Not needed
Fix
None listed

What is CVE-2018-10601?

CVE-2018-10601 is a high-severity memory corruption vulnerability in ICS-CERT IntelliVue Patient Monitors, Avalon Fetal/Maternal Monitors, published on 5 Jun 2018. It can only be exploited from the same machine or local network, with no action from the victim.

Official description

IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that exposes an "echo" service, in which an attacker-sent buffer to an attacker-chosen device address within the same subnet is copied to the stack with no boundary checks, hence resulting in stack overflow.

Weakness: CWE-787, CWE-121

Which versions are affected by CVE-2018-10601?

It affects 18 products.

Products include Philips Avalon Fetal\/Maternal Monitors Fm20 Firmware, Philips Avalon Fetal\/Maternal Monitors Fm30 Firmware, Philips Avalon Fetal\/Maternal Monitors Fm40 Firmware, Philips Avalon Fetal\/Maternal Monitors Fm50 Firmware, Philips Intellivue Mp2 Firmware, Philips Intellivue Mp30 Firmware and 12 more.

How do I fix CVE-2018-10601?

No fix has been published for CVE-2018-10601 yet. Limit exposure of the affected software until one is available.

Upgrade to the fixed version. Memory-safe languages, compiler hardening and fuzzing prevent most new cases. More about Memory corruption.

Is CVE-2018-10601 being exploited?

There is no known exploitation of CVE-2018-10601 in the wild and no public exploit code.

Sources

Also known as GHSA-9f6x-r29g-cv4j.

Is CVE-2018-10601 in your software?

Upload an SBOM and sbomify checks it against every vulnerability here, every day.

Check your SBOMs for free