CVE-2018-8870 Medtronic MyCareLink Patient Monitor Use of Hard-coded Password

Medium 6.4 Published Updated

An attacker with physical access can use a hard-coded password to gain privileged operating system access on Medtronic MyCareLink monitors.

Exploited in the wild
Not known
Login needed
Not needed
Attackable
Local or same network only
Victim action needed
Not needed
Fix
None listed

What is CVE-2018-8870?

CVE-2018-8870 is a medium-severity vulnerability in Medtronic 24950 MyCareLink Monitor, published on 2 Jul 2018. It can only be exploited from the same machine or local network, with no action from the victim.

Official description

Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains a hard-coded operating system password. An attacker with physical access can remove the case of the device, connect to the debug port, and use the password to gain privileged access to the operating system.

Weakness: CWE-798, CWE-259

Which versions are affected by CVE-2018-8870?

It affects 2 products.

Products include Medtronic 24950 Mycarelink Monitor Firmware, Medtronic 24952 Mycarelink Monitor Firmware.

How do I fix CVE-2018-8870?

No fix has been published for CVE-2018-8870 yet. Limit exposure of the affected software until one is available.

Is CVE-2018-8870 being exploited?

There is no known exploitation of CVE-2018-8870 in the wild and no public exploit code.