CVE-2019-13530 Hardcoded credentials in Philips IntelliVue WLAN, portable patient monitors
Hardcoded FTP credentials allow attackers to upload malicious firmware to Philips IntelliVue patient monitors, enabling code execution and privilege escalation.
- Exploited in the wild
- Not known
- Login needed
- An administrator
- Attackable
- Over the internet
- Victim action needed
- Not needed
- Fix
- None listed
What is CVE-2019-13530?
CVE-2019-13530 is a high-severity vulnerability in Philips IntelliVue WLAN, portable patient monitors, published on 12 Sep 2019. An attacker needs administrator access to exploit it, with no action from the victim.
Official description
Philips IntelliVue WLAN, portable patient monitors, WLAN Version A, Firmware A.03.09, WLAN Version A, Firmware A.03.09, Part #: M8096-67501, WLAN Version B, Firmware A.01.09, Part #: N/A (Replaced by Version C) and WLAN Version B, Firmware A.01.09, Part #: N/A (Replaced by Version C). An attacker can use these credentials to login via ftp and upload a malicious firmware.
Which versions are affected by CVE-2019-13530?
It affects 4 products.
Products include Philips Intellivue Mp Monitors Mp20-Mp90 Firmware, Philips Intellivue Mp Monitors Mp2\/X2 Firmware, Philips Intellivue Mp Monitors Mp5\/5sc Firmware, Philips Intellivue Mp Monitors Mx800\/700\/600 Firmware.
How do I fix CVE-2019-13530?
No fix has been published for CVE-2019-13530 yet. Limit exposure of the affected software until one is available.
Is CVE-2019-13530 being exploited?
There is no known exploitation of CVE-2019-13530 in the wild and no public exploit code.
Other vulnerabilities in Philips IntelliVue WLAN, portable patient monitors
Sources
Also known as GHSA-q33g-568m-5mjj.