CVE-2019-25719 Dräger Infinity M540 VG4.1.1 Spoofing and DoS via Network Message Handling
Network-adjacent attackers can spoof data, modify alarm settings, and cause denial-of-service on Dräger Infinity patient monitors.
- Exploited in the wild
- Not known
- Login needed
- Not needed
- Attackable
- Over the internet
- Victim action needed
- Not needed
- Fix
- None listed
What is CVE-2019-25719?
CVE-2019-25719 is a high-severity vulnerability in Dräger Infinity Acute Care System, published on 2 Jun 2026. An attacker can exploit it over the internet without logging in, with no action from the victim.
Official description
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.
Weakness: CWE-924
Which versions are affected by CVE-2019-25719?
How do I fix CVE-2019-25719?
No fix has been published for CVE-2019-25719 yet. Limit exposure of the affected software until one is available.
Is CVE-2019-25719 being exploited?
There is no known exploitation of CVE-2019-25719 in the wild and no public exploit code.