CVE-2019-25722 Dräger SC Monitoring Devices Hard-coded Credentials and DoS
Attackers can use hard-coded credentials to alter device configuration or send malformed packets to cause denial of service in Dräger SC monitoring devices.
- Exploited in the wild
- Not known
- Login needed
- Not needed
- Attackable
- Local or same network only
- Victim action needed
- Not needed
- Fix
- None listed
What is CVE-2019-25722?
CVE-2019-25722 is a high-severity vulnerability in Dräger SC 6002XL, published on 2 Jun 2026. It can only be exploited from the same machine or local network, with no action from the victim.
Official description
Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain hard-coded plaintext credentials in source code and a denial-of-service vulnerability that allows local and remote attackers to compromise device integrity across all software versions. A local attacker with direct device access can use the hard-coded credentials to access service and clinical accounts and alter device configuration, while a remote attacker can send malformed network packets to cause repeated device reboots, ultimately resulting in loss of network connectivity and disruption of patient monitoring.
Weakness: CWE-798
Which versions are affected by CVE-2019-25722?
How do I fix CVE-2019-25722?
No fix has been published for CVE-2019-25722 yet. Limit exposure of the affected software until one is available.
Is CVE-2019-25722 being exploited?
There is no known exploitation of CVE-2019-25722 in the wild and no public exploit code.