CVE-2019-6538 Medtronic Conexus Radio Frequency Telemetry Protocol Improper Access Control

Critical 9.3 Published Updated

Attackers with short-range radio access can intercept, modify, or inject data into Medtronic implanted cardiac devices due to lack of authentication in the telemetry protocol.

Exploited in the wild
Not known
Login needed
Not needed
Attackable
Local or same network only
Victim action needed
Not needed
Fix
None listed

What is CVE-2019-6538?

CVE-2019-6538 is a critical missing authorization vulnerability in Medtronic Amplia CRT-D, published on 25 Mar 2019. It can only be exploited from the same machine or local network, with no action from the victim.

Official description

The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.

Weakness: CWE-862, CWE-306, CWE-284

Which versions are affected by CVE-2019-6538?

It affects 20 products.

Products include Medtronic Amplia Crt-D Firmware, Medtronic Carelink 2090 Firmware, Medtronic Carelink Monitor Firmware, Medtronic Claria Crt-D Firmware, Medtronic Compia Crt-D Firmware, Medtronic Concerto Crt-D Firmware and 14 more.

How do I fix CVE-2019-6538?

No fix has been published for CVE-2019-6538 yet. Limit exposure of the affected software until one is available.

Upgrade, and check permissions on every request on the server, not only in the interface. More about Missing authorization.

Is CVE-2019-6538 being exploited?

There is no known exploitation of CVE-2019-6538 in the wild and no public exploit code.

Sources

Also known as GHSA-56wm-53rq-h4x6.

Is CVE-2019-6538 in your software?

Upload an SBOM and sbomify checks it against every vulnerability here, every day.

Check your SBOMs for free