CVE-2020-12039 Hardcoded credentials in Baxter Sigma Spectrum Infusion Pumps

Low 2.4 Published Updated

Hardcoded passwords allow physical attackers to access biomedical menus, view calibration values, and modify network configurations on Baxter infusion pumps.

Exploited in the wild
Not known
Login needed
Not needed
Attackable
Local or same network only
Victim action needed
Not needed
Fix
None listed

What is CVE-2020-12039?

CVE-2020-12039 is a low-severity vulnerability in Baxter Sigma Spectrum Infusion Pumps, published on 29 Jun 2020. It can only be exploited from the same machine or local network, with no action from the victim.

Official description

Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum Infusion System v's6.x model 35700BAX & Baxter Spectrum Infusion System v's8.x model 35700BAX2 contain hardcoded passwords when physically entered on the keypad provide access to biomedical menus including device settings, view calibration values, network configuration of Sigma Spectrum WBM if installed.

Weakness: CWE-798, CWE-259

Which versions are affected by CVE-2020-12039?

It affects 1 product.

Products include Baxter Sigma Spectrum Infusion System Firmware.

How do I fix CVE-2020-12039?

No fix has been published for CVE-2020-12039 yet. Limit exposure of the affected software until one is available.

Is CVE-2020-12039 being exploited?

There is no known exploitation of CVE-2020-12039 in the wild and no public exploit code.