CVE-2020-15483 Auth bypass in n/a

Medium 6.8 Published Updated

Attackers with physical access can gain root shell access via an unprotected UART debug port on Nescomed Multipara Monitor M1000 devices.

Exploited in the wild
Not known
Login needed
Not needed
Attackable
Local or same network only
Victim action needed
Not needed
Fix
None listed

What is CVE-2020-15483?

CVE-2020-15483 is a medium-severity authentication bypass vulnerability in n/a, published on 26 Aug 2020. It can only be exploited from the same machine or local network, with no action from the victim.

Official description

An issue was discovered on Nescomed Multipara Monitor M1000 devices. The physical UART debug port provides a shell, without requiring a password, with complete access.

Weakness: CWE-306

Which versions are affected by CVE-2020-15483?

It affects 1 product.

Products include Niscomed M1000 Multipara Patient Monitor Firmware.

How do I fix CVE-2020-15483?

No fix has been published for CVE-2020-15483 yet. Limit exposure of the affected software until one is available.

Upgrade, and rotate credentials and sessions that may have been exposed. More about Authentication bypass.

Is CVE-2020-15483 being exploited?

There is no known exploitation of CVE-2020-15483 in the wild and no public exploit code.

Sources

Also known as GHSA-g8hh-gfgv-c2m8.