CVE-2020-25150 B. Braun SpaceCom, Battery Pack SP with Wi-Fi, and Data module compactplus

High 7.6 Published Updated

Attackers with service privileges can exploit path traversal to upload malicious files and execute arbitrary commands on B. Braun medical devices.

Exploited in the wild
Not known
Login needed
Any signed-in user
Attackable
Over the internet
Victim action needed
Not needed
Fix
None listed

What is CVE-2020-25150?

CVE-2020-25150 is a high-severity path traversal vulnerability in B. Braun Melsungen AG Battery pack with Wi-Fi, published on 14 Apr 2022. Any signed-in user can exploit it over the network, with no action from the victim.

Official description

A relative path traversal attack in the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers with service user privileges to upload arbitrary files. By uploading a specially crafted tar file an attacker can execute arbitrary commands.

Weakness: CWE-22, CWE-23

Which versions are affected by CVE-2020-25150?

It affects 2 products.

Products include Bbraun Datamodule Compactplus, Bbraun Spacecom.

How do I fix CVE-2020-25150?

No fix has been published for CVE-2020-25150 yet. Limit exposure of the affected software until one is available.

Upgrade, and resolve every user-supplied path against an allowed base folder before opening it. More about Path traversal.

Is CVE-2020-25150 being exploited?

There is no known exploitation of CVE-2020-25150 in the wild and no public exploit code.