CVE-2021-23195 Fresenius Kabi Agilia Connect Infusion System exposure of information through directory listing
Attackers can view server directory contents and access files if automated indexing is enabled in the Vigilant Software Suite dashboard.
- Exploited in the wild
- Not known
- Login needed
- Not needed
- Attackable
- Over the internet
- Victim action needed
- Not needed
- Fix
- None listed
What is CVE-2021-23195?
CVE-2021-23195 is a medium-severity vulnerability in Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard), published on 21 Jan 2022. An attacker can exploit it over the internet without logging in, with no action from the victim.
Official description
Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 has the option for automated indexing (directory listing) activated. When accessing a directory, a web server delivers its entire content in HTML form. If an index file does not exist and directory listing is enabled, all content of the directory will be displayed, allowing an attacker to identify and access files on the server.
Which versions are affected by CVE-2021-23195?
It affects 6 products.
Products include Fresenius-Kabi Agilia Connect Firmware, Fresenius-Kabi Agilia Partner Maintenance Software, Fresenius-Kabi Link\+ Agilia Firmware, Fresenius-Kabi Vigilant Centerium, Fresenius-Kabi Vigilant Insight, Fresenius-Kabi Vigilant Mastermed.
How do I fix CVE-2021-23195?
No fix has been published for CVE-2021-23195 yet. Limit exposure of the affected software until one is available.
Is CVE-2021-23195 being exploited?
There is no known exploitation of CVE-2021-23195 in the wild and no public exploit code.
Other vulnerabilities in Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard)
- CVE-2021-23236 Fresenius Kabi Agilia Connect Infusion System uncontrolled resource consumption
- CVE-2021-23196 Fresenius Kabi Agilia Connect Infusion System insufficiently protected credentials
- CVE-2021-23233 Fresenius Kabi Agilia Connect Infusion System
- CVE-2021-43355 Fresenius Kabi Agilia Connect Infusion System use of client side authentication
- CVE-2021-31562 Fresenius Kabi Agilia Connect Infusion System use of a broken or risky cryptographic algorithm
Sources
Also known as GHSA-29g2-4fxx-7xmh.