CVE-2021-23233 Fresenius Kabi Agilia Connect Infusion System
Unauthenticated attackers can access sensitive endpoints to perform critical actions or modify configuration parameters in the Agilia Link+ infusion system.
- Exploited in the wild
- Not known
- Login needed
- Not needed
- Attackable
- Over the internet
- Victim action needed
- Not needed
- Fix
- None listed
What is CVE-2021-23233?
CVE-2021-23233 is a high-severity missing authorization vulnerability in Fresenius Kabi Agilia Link+, published on 21 Jan 2022. An attacker can exploit it over the internet without logging in, with no action from the victim.
Official description
Sensitive endpoints in Fresenius Kabi Agilia Link+ v3.0 and prior can be accessed without any authentication information such as the session cookie. An attacker can send requests to sensitive endpoints as an unauthenticated user to perform critical actions or modify critical configuration parameters.
Which versions are affected by CVE-2021-23233?
It affects 6 products.
Products include Fresenius-Kabi Agilia Connect Firmware, Fresenius-Kabi Agilia Partner Maintenance Software, Fresenius-Kabi Link\+ Agilia Firmware, Fresenius-Kabi Vigilant Centerium, Fresenius-Kabi Vigilant Insight, Fresenius-Kabi Vigilant Mastermed.
How do I fix CVE-2021-23233?
No fix has been published for CVE-2021-23233 yet. Limit exposure of the affected software until one is available.
Upgrade, and check permissions on every request on the server, not only in the interface. More about Missing authorization.
Is CVE-2021-23233 being exploited?
There is no known exploitation of CVE-2021-23233 in the wild and no public exploit code.
Other vulnerabilities in Fresenius Kabi Agilia Link+
- CVE-2021-23236 Fresenius Kabi Agilia Connect Infusion System uncontrolled resource consumption
- CVE-2021-23196 Fresenius Kabi Agilia Connect Infusion System insufficiently protected credentials
- CVE-2021-43355 Fresenius Kabi Agilia Connect Infusion System use of client side authentication
- CVE-2021-31562 Fresenius Kabi Agilia Connect Infusion System use of a broken or risky cryptographic algorithm
- CVE-2021-44464 Fresenius Kabi Agilia Connect Infusion System hard coded credentials
Sources
Also known as GHSA-2vv3-q2gp-2rjv.