CVE-2021-37165 Memory corruption in n/a

Critical 9.8 Published Updated

A buffer overflow in the HMI Control Panel allows remote code execution via crafted TCP messages, affecting healthcare infrastructure.

Exploited in the wild
Not known
Login needed
Not needed
Attackable
Over the internet
Victim action needed
Not needed
Fix
None listed

What is CVE-2021-37165?

CVE-2021-37165 is a critical memory corruption vulnerability in n/a, published on 2 Aug 2021. An attacker can exploit it over the internet without logging in, with no action from the victim.

Official description

A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. When a message is sent to the HMI TCP socket, it is forwarded to the hmiProcessMsg function through the pendingQ, and may lead to remote code execution.

Weakness: CWE-120

Which versions are affected by CVE-2021-37165?

It affects 1 product.

Products include Swisslog-Healthcare Hmi-3 Control Panel Firmware.

How do I fix CVE-2021-37165?

No fix has been published for CVE-2021-37165 yet. Limit exposure of the affected software until one is available.

Upgrade to the fixed version. Memory-safe languages, compiler hardening and fuzzing prevent most new cases. More about Memory corruption.

Is CVE-2021-37165 being exploited?

There is no known exploitation of CVE-2021-37165 in the wild and no public exploit code.

Sources

Also known as GHSA-jfh4-jcmh-hq55.