CVE-2021-38394 Security feature bypass in ZOOM LATITUDE
An attacker with physical access can reverse engineer the hardware key check to create a duplicate key, bypassing access controls for special settings.
- Exploited in the wild
- Not known
- Login needed
- Not needed
- Attackable
- Local or same network only
- Victim action needed
- A user has to click or open something
- Fix
- None listed
What is CVE-2021-38394?
CVE-2021-38394 is a medium-severity vulnerability in Boston Scientific ZOOM LATITUDE, published on 4 Oct 2021. It can only be exploited from the same machine or local network, and only if a user opens a file or follows a link.
Official description
An attacker with physical access to the device can extract the binary that checks for the hardware key and reverse engineer it, which could be used to create a physical duplicate of a valid hardware key. The hardware key allows access to special settings when inserted.
Weakness: CWE-1278
Which versions are affected by CVE-2021-38394?
It affects 1 product.
Products include Bostonscientific Zoom Latitude Pogrammer\/Recorder\/Monitor 3120 Firmware.
How do I fix CVE-2021-38394?
No fix has been published for CVE-2021-38394 yet. Limit exposure of the affected software until one is available.
Is CVE-2021-38394 being exploited?
There is no known exploitation of CVE-2021-38394 in the wild and no public exploit code.
Other vulnerabilities in Boston Scientific ZOOM LATITUDE
- CVE-2021-38400 Use of Password Hash with Insufficient Computational Effort for Boston Scientific Zoom Latitude
- CVE-2021-38392 Improper Access Control for Boston Scientific Zoom Latitude
- CVE-2021-38396 Missing Support Integrity Check for Boston Scientific Zoom Latitude
- CVE-2021-38398 Reliance on Component that is not Updateable for Boston Scientific Zoom Latitude
Sources
Also known as GHSA-c3hf-72hw-fvr7.