CVE-2024-1630 Path traversal in LOGIQ E10
An attacker can exploit a path traversal flaw in GE HealthCare ultrasound devices to read arbitrary files from the system.
- Exploited in the wild
- Not known
- Login needed
- Not needed
- Attackable
- Local or same network only
- Victim action needed
- Not needed
- Fix
- None listed
What is CVE-2024-1630?
CVE-2024-1630 is a high-severity path traversal vulnerability in GE HealthCare LOGIQ E10, published on 14 May 2024. It can only be exploited from the same machine or local network, with no action from the victim.
Official description
Path traversal vulnerability in “getAllFolderContents” function of Common Service Desktop, a GE HealthCare ultrasound device component
Weakness: CWE-22
Which versions are affected by CVE-2024-1630?
How do I fix CVE-2024-1630?
No fix has been published for CVE-2024-1630 yet. Limit exposure of the affected software until one is available.
Upgrade, and resolve every user-supplied path against an allowed base folder before opening it. More about Path traversal.
Is CVE-2024-1630 being exploited?
It is not confirmed as exploited in the wild, but public exploit code for CVE-2024-1630 exists.
Sources
Also known as GHSA-gwj2-vqgr-4qxf.