CVE-2024-48973 Debug port on Life2000 Ventilator serial interface is enabled by default

Critical 9.3 Published Updated

Attackers can access unencrypted debug ports on Life2000 ventilators to disclose information or alter device settings due to default configurations.

Exploited in the wild
Not known
Login needed
Not needed
Attackable
Local or same network only
Victim action needed
Not needed
Fix
None listed

What is CVE-2024-48973?

CVE-2024-48973 is a critical vulnerability in Baxter Life2000 Ventilation System, published on 14 Nov 2024. It can only be exploited from the same machine or local network, with no action from the victim.

Official description

The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

Weakness: CWE-1263

Which versions are affected by CVE-2024-48973?

How do I fix CVE-2024-48973?

No fix has been published for CVE-2024-48973 yet. Limit exposure of the affected software until one is available.

Is CVE-2024-48973 being exploited?

There is no known exploitation of CVE-2024-48973 in the wild and no public exploit code.

Sources

Also known as GHSA-xx59-963p-2v9w.

Is CVE-2024-48973 in your software?

Upload an SBOM and sbomify checks it against every vulnerability here, every day.