CVE-2024-48974 Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates

Critical 9.3 Published Updated

Attackers can push compromised firmware to the Life2000 Ventilator due to missing integrity checks, potentially disrupting device function or disclosing information.

Exploited in the wild
Not known
Login needed
Not needed
Attackable
Local or same network only
Victim action needed
Not needed
Fix
None listed

What is CVE-2024-48974?

CVE-2024-48974 is a critical vulnerability in Baxter Life2000 Ventilation System, published on 14 Nov 2024. It can only be exploited from the same machine or local network, with no action from the victim.

Official description

The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.

Weakness: CWE-494

Which versions are affected by CVE-2024-48974?

How do I fix CVE-2024-48974?

No fix has been published for CVE-2024-48974 yet. Limit exposure of the affected software until one is available.

Is CVE-2024-48974 being exploited?

There is no known exploitation of CVE-2024-48974 in the wild and no public exploit code.

Sources

Also known as GHSA-rfvx-872w-6rpr.

Is CVE-2024-48974 in your software?

Upload an SBOM and sbomify checks it against every vulnerability here, every day.