CVE-2025-1204 Hardcoded credentials in CMS8000 Patient Monitor

High 7.7 Published Updated

Attackers can overwrite device files by impersonating a hard-coded IP address during boot, potentially leading to code execution on the patient monitor.

Exploited in the wild
Not known
Login needed
Not needed
Attackable
Over the internet
Victim action needed
A user has to click or open something
Fix
None listed

What is CVE-2025-1204?

CVE-2025-1204 is a high-severity vulnerability in Contec Health CMS8000 Patient Monitor, published on 25 Feb 2025. An attacker can exploit it over the internet without logging in, and only if a user opens a file or follows a link.

Official description

The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the 'C' button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device.

Weakness: CWE-912

Which versions are affected by CVE-2025-1204?

How do I fix CVE-2025-1204?

No fix has been published for CVE-2025-1204 yet. Limit exposure of the affected software until one is available.

Is CVE-2025-1204 being exploited?

It is not confirmed as exploited in the wild, but public exploit code for CVE-2025-1204 exists.