CVE-2025-14096 Credential Disclosure vulnerability in Radiometer Products
Attackers with physical access can extract credentials from Radiometer medical analyzers due to insufficient OS protection, requiring local vendor contact for resolution.
- Exploited in the wild
- Not known
- Login needed
- Not needed
- Attackable
- Local or same network only
- Victim action needed
- Not needed
- Fix
- None listed
What is CVE-2025-14096?
CVE-2025-14096 is a high-severity vulnerability in Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX analyzers, published on 17 Dec 2025. It can only be exploited from the same machine or local network, with no action from the victim.
Official description
A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.
Other related CVE's are CVE-2025-14095 & CVE-2025-14097.
Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.
Required Configuration for Exposure:
Attacker requires physical access to the analyzer.
Temporary work Around: Only authorized people can physically access the analyzer.
Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution.
Exploit Status:
Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.
Which versions are affected by CVE-2025-14096?
How do I fix CVE-2025-14096?
No fix has been published for CVE-2025-14096 yet. Limit exposure of the affected software until one is available.
Is CVE-2025-14096 being exploited?
There is no known exploitation of CVE-2025-14096 in the wild and no public exploit code.
Sources
Also known as GHSA-h5qw-42jq-7cmh.