CVE-2025-14097 Remote Code Execution Vulnerability in Radiometer Products

High 7.2 Published Updated

Attackers can achieve remote code execution and unauthorized device management on Radiometer analyzers if the remote support feature is enabled.

Exploited in the wild
Not known
Login needed
An administrator
Attackable
Over the internet
Victim action needed
Not needed
Fix
None listed

What is CVE-2025-14097?

CVE-2025-14097 is a high-severity authentication bypass vulnerability in Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers, published on 17 Dec 2025. An attacker needs administrator access to exploit it, with no action from the victim.

Official description

A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software.                                                                                                                                                                                                Other related CVE's are CVE-2025-14095 & CVE-2025-14096.                                                                                                      Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.

Required Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer.                                                                                                                                                                        Temporary work Around: If the network is not considered secure, please remove the analyzer from the network.                         Permanent solution: Customers should ensure the following: • The network is secure, and access follows best practices. Local Radiometer representatives will contact all affected customers to discuss a permanent solution.                                                      Exploit Status:

Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.

Weakness: CWE-287

Which versions are affected by CVE-2025-14097?

How do I fix CVE-2025-14097?

No fix has been published for CVE-2025-14097 yet. Limit exposure of the affected software until one is available.

Upgrade, and rotate credentials and sessions that may have been exposed. More about Authentication bypass.

Is CVE-2025-14097 being exploited?

There is no known exploitation of CVE-2025-14097 in the wild and no public exploit code.

Sources

Also known as GHSA-rjfq-mp68-v4f7.

Is CVE-2025-14097 in your software?

Upload an SBOM and sbomify checks it against every vulnerability here, every day.

Check your SBOMs for free