CVE-2025-62575 Mirion Medical EC2 Software NMIS BioDose Incorrect Permission Assignment for Critical Resource

High 8.7 Published Updated

Default SQL user accounts have sysadmin privileges, allowing remote code execution via stored procedures in Mirion Medical dosimetry software.

Exploited in the wild
Not known
Login needed
Any signed-in user
Attackable
Over the internet
Victim action needed
Not needed
Fix
None listed

What is CVE-2025-62575?

CVE-2025-62575 is a high-severity vulnerability in Mirion Medical EC2 Software NMIS BioDose, published on 2 Dec 2025. Any signed-in user can exploit it over the network, with no action from the victim.

Official description

NMIS/BioDose V22.02 and previous versions rely on a Microsoft SQL Server database. The SQL user account 'nmdbuser' and other created accounts by default have the sysadmin role. This can lead to remote code execution through the use of certain built-in stored procedures.

Weakness: CWE-732

Which versions are affected by CVE-2025-62575?

It affects 1 product.

Products include Mirion Biodose\/Nmis.

How do I fix CVE-2025-62575?

No fix has been published for CVE-2025-62575 yet. Limit exposure of the affected software until one is available.

Is CVE-2025-62575 being exploited?

There is no known exploitation of CVE-2025-62575 in the wild and no public exploit code.

Sources

Also known as GHSA-959v-frv4-7fj8.

Is CVE-2025-62575 in your software?

Upload an SBOM and sbomify checks it against every vulnerability here, every day.